3.2 Tutorial

711 Liquino, Instructions for use

54

SEQ method editing


As the sub-methods have now been defined and stored, the SEQ method
can now be assembled.

*

MODE

-

X

DOS NR

SELECT

P

The SEQ mode must first be activated. This sets all method parameters

to standard values. Press the key and then


The key is used to access the parameter menu, where all

the possible settings for a SEQ method can be found.

Di spl ay:

par amet er s

>SEQ sequence

In the submenu ‘SEQ sequence’ previously defined and stored XDOS

methods can be assembled to form a sequence and the correspond-
ing pauses inserted. Press to open the submenu.

Di spl ay:

>SEQ sequence st ep 1

command: NO OPERATI ON

"1. Temperature-controlled acidification of aniline with semi-

concentrated hydrochloric acid (temperature below approx. 10 °C)“.

Use the key until the name of the method which has
been stored for the acidification appears. Press to accept
the selection.

Di spl ay:

>SEQ sequence st ep 2

command: NO OPERATI ON

"2. Pause for 5 minutes for cooling down to 0 °C.“

The

Di spl ay:

>SEQ sequence st ep 3

command: NO OPERATI ON

"3. Temperature-controlled diazotization (temperature below 5 °C).“

Select the command ‘METHOD’ as in step 1 and then the XDOS meth-
od which has been stored for diazotization. Confirm each with
.

Di spl ay:

>SEQ sequence st ep 4