3 Introduction

711 Liquino, Instructions for use

31








dr i ve uni t : DOS1

The Dosino which is to be used can be selected here with until the dosing unit defined in the above
preparation step appears. (If the pipetting attachment is used this set-
ting can be left as ‘standard’). Confirm with .

Di spl ay:

par amet er s

>PI P r at es

As the ‘PIP dosing unit’ submenu contains no further menu items, an

automatic return is made to the submenu selection. All necessary pa-
rameters have now been entered! The parameter menu is exited with
.

Di spl ay:

PI P * * * * * * * *

r eady 0. 000 mL



If an inquiry is always to be made about the pipetting volume at the start of
the method then enter the value under ‘parameters/>PIP preselec-
tions/query pip. vol.’.

Various PIP methods can also be edited and then stored in the method
memory. Each of these methods can then be reloaded by pressing a few
keys. The method memory is described on page 128.


Carrying out the method

START

Z

When is now pressed the display shows

Di spl ay:

Not pr epar ed yet !

Run PREP now?

ENTER

YES




<1> , <2>

Before the Dosino is used for the first time in the new configuration the

Dosino cylinder and the complete tubing system must be filled bubble-
free with reagent solution. Press .

Di spl ay:

wast e por t :

t i p( 1) or f l ask( 2) ?

Press <1> to empty the cylinder contents through the dosing tip, or

<2> to return them to the bottle. Take care that the tip of the dosing
tubing is located in a vessel. Wait until the 711 Liquino has ended the
preparation (PREP). The instrument then returns to its normal state.