3 cont mode parameters – Metrohm 711 Liquino User Manual
Page 102
4.6 CONT mode - preparation of solutions
711 Liquino, Instructions for use
96
4.6.3
CONT mode parameters
PARAM
B
All settings in the parameter menu form a method and can be stored as
such.
Main menu:
par amet er s
>CONT par amet er
Open submenu with
par amet er s
>CONT dosi ng uni t s
Use <> or <> to move down or up from one menu point to the
next
par amet er s
>CONT r at es
Use
par amet er s
>CONT pr esel ect i ons
Return to basic condition with
par amet er s
>CONT r epor t s
Submenus:
par amet er s
>CONT par amet er
Dosing parameter settings
Navigate with
>CONT par amet er
t ype: subst . conc
Select the dosing procedure
subst.conc.
mass conc..
mass fraction
volume fraction
= solutions in mol/L
= solutions in g/L
= solutions in %
= dilutions in mol/L, ppm, g/L, %
>CONT par amet er
t ar get cont ent 1. 0 Mol / L
Target concentration
0.001…9999.9
The unit depends on the 'type',
see above.
>CONT par amet er
t ar get vol ume 10. 0 mL
Target volume
0.01…9999
.
>CONT par amet er
f act or 1. 0
Correction factor
0.0000001…999999.0
not for 'volume fraction'.
>CONT par amet er
mol ar mass 1. 0 g/ Mol
Molar mass of the substance
0.0001…999999
only for 'subst.conc'.
>CONT par amet er
densi t y 1. 0 g/ mL
Density of solvent
0.1…99.9999
only for 'mass fraction'.
>CONT par amet er
meas. uni t Mol / L
Mass unit of target content
mol/L, ppm, g/L, %
only for 'volume fraction'.