Metrohm 711 Liquino User Manual
Page 31
3 Introduction
711 Liquino, Instructions for use
25
•
If you do not want the dosing speed to increase or you require a quicker
increase in the dosing speed then you can set this in the configuration.
CONFIG
A
2x
J
•
Press the
Di spl ay:
conf i gur at i on
>manual
•
Here you will find the submenu with the settings which are effective for
manual functions.
Di spl ay:
>manual DOS1
← dosing connection
dos. uni t : def aul t
•
In the first line, above right, the address of the dosing connection to
which the particular setting applies is shown. You can make different
settings for each of the four dosing connections. The settings for the
other dosing connections DOS2 to DOS4 are also found in this sub-
menu. Now press the
Di spl ay:
>manual DOS1
>>dosi ng
•
If you want to alter the dosing behavior of dosing device 1 you will find
this in the second submenu '>>Dosing' after pressing the
key as given above. If your Dosino is connected to 'Dos. 2' then press
the
Di spl ay:
>>dosi ng DOS1
r at e t ype: i ncr easi ng
•
By pressing the
can switch to a constant dosing speed. You should, however, leave the
setting with increasing speed alone and press
Di spl ay:
>>dosi ng DOS1
sl ope 4
•
Enter the highest value '10' for the slope of the increase of the dosing
speed. Confirm the entry with
•
By pressing the
ing.
Di spl ay:
manual DOSI NG
r eady 3. 349 mL
START
Z
•
Start dosing again by pressing the
dosing speed increases more rapidly and that the highest dosing speed is
reached very quickly.
FILL
5
M
•
When the max. cylinder volume has been reached the dosing unit will be
filled automatically. Filling the dosing cylinder can also be triggered manu-
ally by pressing the