4.8 SMPL mode - fully automatic sampling
711 Liquino, Instructions for use
116
>SMPL dosi ng uni t s
>>aux. sol ut . AFTER
Submenu for definition of dosing proce-
dures after sample transfer
>>aux. sol ut . AFTER no1
dosi ng dr i ve: of f
Dosing connection for addition of the first
auxiliary solution
…up to 3 auxiliary dosing devices
See
>>aux.solut. BEFORE
par amet er s
>SMPL cont r ol I / O
Settings for controlling external instruments
Navigate with
>SMPL cont r ol I / O
sampl e changer ct r l : of f
Use of a sample changer
on, off
with 'on' the following settings can be made:
>>sampl e changer cont r ol
st ar t vi a: r emot e
Type of connection to sample changer
Remote,
RS
= connection to parallel remote socket
= connection to serial RS232 interface
>>sampl e changer cont r ol
st ar t cmd &M; $G
Remote control command to start the
sample changer via RS232
11 ASCII characters
only for 'start via RS'
see description of Metrohm remote control lan-
guage in the
'Remote control reference'
>>sampl e changer cont r ol
awai t acknowl edge: of f
Type of sample changer acknowledgment
(ready message)
Remote,
RS,
off
= acknowledgment via remote socket
= acknowledgment via RS232 interface
= no acknowledgment
>>sampl e changer cont r ol
ack. st r i ng *
Sample changer acknowledgment as
character string via RS232
11 ASCII characters
only for 'Start via RS'
* = any acknowledgment
see description of Metrohm remote control lan-
guage in the
'Remote control reference'
>SMPL cont r ol I / O
st ar t ext . devi ce: of f
Start of external instrument (e.g. Titrino)to de-
termine the sample
on,
off
with 'on' the following settings can be made:
>>ext . devi ce cont r ol
st ar t vi a: r emot e
Type of connection to external instrument
remote,
RS
= connection to parallel remote socket
= connection to serial RS232 interface
>>ext . devi ce cont r ol
st ar t cmd &M; $G
Remote control command to start the
external instrument via RS232
11 ASCII characters
only for 'Start via RS'
see description of Metrohm remote control lan-
guage in the
'Remote control reference'