2 smpl mode parameters – Metrohm 711 Liquino User Manual
Page 119
![background image](/manuals/587118/119/background.png)
4. Detailed description
711 Liquino, Instructions for use
113
4.8.2
SMPL mode parameters
PARAM
B
All settings in the parameter menu form a method and can be stored as
such.
Main menu:
par amet er s
>SMPL par amet er
Open submenu with
par amet er s
>SMPL sampl i ng oper at i on
Use <> or <> to move down or up from one menu point to the
next
par amet er s
>SMPL meas. cel l r i nsi ng
Use
par amet er s
>SMPL dosi ng uni t s
Return to basic condition with
par amet er s
>SMPL cont r ol I / O
par amet er s
>SMPL pr esel ect i ons
par amet er s
>SMPL r epor t s
Submenus:
par amet er s
>SMPL par amet er
Process parameter settings
Navigate with
>SMPL par amet er
sampl e count 1
Number of samples
1…99999,
infinite
= continuous cycle, select with
>SMPL par amet er
sampl i ng mode: manual
Type of sampling
manual,
auto
= sample changed by hand (with separating bubble)
= sampling with sample changer or series with the same
sample (multiple determinations)
>SMPL par amet er
t i me i nt er val 00: 00: 00
Waiting time between individual samples
HH:MM:SS
par amet er s
>SMPL sampl i ng oper at i on
Setting the sampling parameters
Navigate with
>SMPL sampl i ng oper at i on
sampl e vol ume 10 mL
Sample volume
0.001…50.0
>SMPL sampl i ng oper at i on
di l ut i on vol . 10 mL
Solvent volume for dilution of sample
0.001…50.0
>SMPL sampl i ng oper at i on
sampl e r i nse vol . 10 mL
Sample volume for pre-rinsing the dosing
cylinder
0.001…50.0 recommended: 20…50% of the sample volume