5.2 Please note …
711 Liquino, Instructions for use
136
chloromethane, trinitrophenol-2,3,6, vinyl acetate, vinyl chloride, vinyli-
dene chloride
Never dose
hot (>60°C) chemicals or solutions with Dosinos!
•
Metrohm offers tubing made of
FEP (tetrafluoroethylene-
hexafluoropropylene copolymer) and
PTFE (polytetrafluoroethylene).
These materials are largely resistant to organic solvents, strong acids
and alkalis.
FEP tubing is not suitable for:
−
perchloric acid
−
fluorine
FEP tubing has a limited resistance to:
−
chlorosulfonic acid
FEP tubing has a limited resistance at higher temperatures to:
−
crude oil products
−
perchloroethylene
−
tetrachloroethylene
PTFE tubing has a limited resistance to:
−
halogenated hydrocarbons
−
perchloric acid
−
fluorine
PTFE tubing has a limited resistance at higher temperatures to:
−
crude oil products
−
perchloroethylene
−
tetrachloroethylene
•
Unoccupied dosing unit connections (Ports) can be sealed with
drying tubes or stoppers.
•
Divide your synthesis run into part steps and create separate XDOS
methods for the individual dosing steps. Then you can link together the
individual part steps (methods) to form a sequence (SEQ mode) and
add further commands to it.