Metrohm 711 Liquino User Manual
Page 123

4. Detailed description
711 Liquino, Instructions for use
117
>>ext . devi ce cont r ol
awai t acknowl edge: of f
Type of acknowledgment from external
instrument (ready message)
Remote,
RS,
off
= acknowledgment via remote socket
= acknowledgment via RS232 interface
= no acknowledgment
>>ext . devi ce cont r ol
ack. st r i ng *
Acknowledgment from external instrument
as character string via RS232
11 ASCII characters
only for 'Start via RS'
* = any acknowledgment
see description of Metrohm remote control lan-
guage in the
'Remote control reference'
par amet er s
>SMPL pr esel ect i ons
Definition of requests at start of sampling
Navigate with
>SMPL pr esel ect i ons
r eq. i dent : of f
Automatic request for up to
3 identification texts
id1,
id1&2,
all,
off
= only ID1 is requested
= ID1 and ID2 are requested
= ID1, ID2 and ID3 are requested
= no request
>SMPL pr esel ect i ons
>>SMPL pr ompt s
Submenu for definition of the
request texts
>>SMPL pr ompt s
i d1 I d1
Display text for requesting the first ID
14 ASCII characters
>>SMPL pr ompt s
i d2 I d2
Display text for requesting the second ID
14 ASCII characters
>>SMPL pr ompt s
i d3 I d3
Display text for requesting the third ID
14 ASCII characters
Use identification texts (ID1…3) for identifying your sample or your test.
The IDs will be printed out in the report of the dosing procedure so that
the individual reports are identified unambiguously.
par amet er s
>SMPL r epor t s
Report definitions for the result output
Navigate with
>SMPL r epor t s no. 1
r epor t : of f
Selection of one or more reports to be printed
out when sampling is completed.
off
result,
parameter,
ff
= no report
= list of relevant results.
= list of all method parameters
= FORM FEED; triggers paper feed at printer
(important for laser and inkjet printers for
ejecting pages)
Up to three reports are possible.