beautypg.com

Metrohm 711 Liquino User Manual

Page 41

background image

3 Introduction

711 Liquino, Instructions for use

35

The dosing units which correspond to the tubing used for the Dosinos must
be defined. This process is described on page 16.


Method editing

*

MODE

-

X

DOS NR

SELECT

P

The SMPL mode must first be activated. This sets all method parame-

ters to standard values. Press the key and then can be used to choose the entry ‘auto’. However, if a dif-
ferent sample is to be transferred each time then ‘manual’ should be
selected. In this case the process requires the sample tubing to be
immersed in the new sample each time. Press .

Di spl ay:

>SMPL par amet er s

t i me i nt er val 00: 05: 00

This parameter determines the time interval between successive sam-

ples. For a multiple determination this value should be set to 00:00:00.
In this way the next sample will be transferred at the end of the titration.
Press to accept the value.

Di spl ay:

par amet er s

>SMPL sampl i ng oper at i on

As the ‘SMPL parameters’ submenu contains no further menu items,

an automatic return is made to the submenu selection. The submenu
‘SMPL sampling operation' contains the parameters for sample trans-
fer. Press to open it.

Di spl ay:

>SMPL sampl i ng oper at i on

sampl e vol ume 3 mL

Enter the volume of the sample to be transferred to the measuring cell

(in this case 3 mL). Press to accept the value.

Di spl ay:

>SMPL sampl i ng oper at i on