4. Detailed description
711 Liquino, Instructions for use
109
Example: remote connection:
Titroprocessor 726
Liquino (under Parameter
>SMPL cont r ol I / O
>>ext . devi ce cont r ol
)
CTRL_RM Remot e A pat t er n * * * * * * * 1 The 711 Liquino is started.
SCAN_RM Remot e A pat t er n * 1* * * * * * st ar t vi a: r emot e
determination (e.g.
DET_PH* )
CTRL_RM Remot e A pat t er n * * * * 1* * * awai t acknowl edge: r emot e
With an RS232 connection any control sequences can be selected. See be-
low.
Example: RS232 connection:
Titroprocessor 726
Liquino (under Parameter
>SMPL cont r ol I / O
>>ext . devi ce cont r ol
)
CTRL_RS1 &M; $G
The 711 Liquino is started.
SCAN_RS1 START
(exam-
ple)
st ar t vi a: RS
st ar t cmd START
(example)
determination (e.g.
DET_PH* )
CTRL_RS1 r eady
(exam-
ple)
awai t acknowl edge: RS
ack. st r i ng r eady
(example)
Sample changer
If a Metrohm 730 Sample Changer is used the possibilities of the 711 Liqui-
no can be extended even further. After the start of the SMPL mode a 730
Sample Changer can be started via a remote lead or a serial RS232 connec-
tion. This requires a special sample changer method which is optimized for
communication with the 711 Liquino. Details for programming sample
changer methods and instrument communication can be found in ‘730
Sample Changer, Instructions for use’.
Methods
All settings in the parameter menu can be stored as a method and reloaded
when required.