4. Detailed description
711 Liquino, Instructions for use
77
4.4.1
XDOS mode parameters
PARAM
B
All settings in the parameter menu form a method and can be stored as
such.
Main menu:
par amet er s
>XDOS par amet er
Open submenu with
par amet er s
>XDOS dosi ng uni t s
Use <> or <> to move down or up from one menu point to the
next
par amet er s
>XDOS moni t or i ng
Use
par amet er s
>XDOS pr esel ect i ons
Return to basic condition with
par amet er s
>XDOS r epor t s
Submenus:
par amet er s
>XDOS par amet er
Dosing parameter settings.
Navigate with
>XDOS par amet er
XDOS t ype: vol -
ume&r at e
Selection of task parameters
volume&rate,
volume&time,
time&rate
= selection of volume and dosing rate
= selection of volume and dosing time
= selection of dosing time and dosing rate
>XDOS par amet er
vol ume 1. 0 mL
Dosing volume
0.0…99999.9,
infinite
not for time&rate
= continuous dosing; can be ended with
>XDOS par amet er
f eed r at e max mL/ mi n
Dosing rate
0.001…166.66,
max
not for volume&time
= highest possible dosing rate (depends on type of
dosing unit.)
The dosing rate is the adjustable piston stroke speed when discharging the
cylinder contents. For simple dosing it does not correspond to the effective
delivery rate as the filling times of the dosing unit must be taken into ac-
count.
>XDOS par amet er
dosi ng t i me 00: 01: 00
Time selected for complete dosing process.
Filing times must be taken into account.
HH:MM:SS
not for volume&rate
in hours, minutes and seconds (corrections can be made
with
>XDOS par amet er
t emp. measur i ng: of f
Temperature measurement (compulsory for
temperature-controlled dosing)
on, off
>XDOS par amet er
r ec. i nt er val 00: 01: 00
Recording interval for
temperature measurement
HH:MM:SS
(only for temp. measuring 'on'
)
in hours, minutes and seconds
Depending on the
selected XDOS type
two of three dosing
parameters will
be shown.
The third parameter will
be automatically
determined by the
Liquino 711 or results
from the dosing itself
and will be printed out in
the result report when
dosing has been con-
cluded.