4. Detailed description
711 Liquino, Instructions for use
123
4.9.1
GLP mode parameters
PARAM
B
All settings in the parameter menu form a method and can be stored as
such.
Main menu:
par amet er s
>GLP par amet er s
Open submenu with
par amet er s
>GLP opt i ons
Use <> or <> to move down or up from one menu point to the
next
par amet er s
>GLP r epor t s
Use
Return to normal state with
Submenus:
par amet er s
>GLP par amet er s
Dosing parameter settings
Navigate with
>GLP par amet er s
dosi ng dr i ve: DOS1
Dosing connection
DOS1, DOS2, DOS3, DOS4
>GLP par amet er s
dos. uni t : def aul t
Name of dosing unit used
selection list see configuration '
>dosi ng uni t s'
>GLP par amet er s
t emper at ur e 20. 00° C
Temperature of liquid
-70…200.0
par amet er s
>GLP opt i ons
Other settings
Navigate with
>GLP opt i ons
l i qui d: wat er
Type of liquid
water, other
>GLP opt i ons
l i qu. name
Name of liquid
13 ASCII characters
(only for 'liquid: other')
Text input with
>GLP opt i ons
l i qu. dens. 1. 0000 g/ mL
Density of liquid at the measuring tem-
perature
0.0100…20.0000
(only for 'liquid: other')
>GLP opt i ons
f eed r at e max mL/ mi n
Dosing rate
0.001…166.66,
max
(only for 'liquid: other')
= highest possible dosing rate (depends on type
of dosing unit.)