Metrohm 711 Liquino User Manual
Page 64
![background image](/manuals/587118/64/background.png)
3.2 Tutorial
711 Liquino, Instructions for use
58
Creating the GLP method
*
MODE
-
X
DOS NR
SELECT
P
•
The GLP mode must first be activated. This sets all method parameters
to standard values. Press the
Di spl ay:
mode sel ect : GLP
ENTER
YES
•
Press
Di spl ay:
GLP * * * * * * * *
r eady 0. 000 mL
PARAM
B
•
The key is used to access the parameter menu, where all
the possible settings for a GLP method can be found.
Di spl ay:
par amet er s
>GLP par amet er s
•
The submenu ‘GLP parameters’ contains the main settings for a GLP
method. Press
Di spl ay:
>GLP par amet er s
dr i ve uni t : DOS1
•
The Dosino which is to be used can be chosen here with
(DOSx refers to the connection number on the 711 Liquino). Press
Di spl ay:
>GLP par amet er s
dos. uni t : def aul t
•
In this menu item the 711 Liquino is informed about the tubing of the
Dosino. Press
Di spl ay:
>GLP par amet er s
t emper at ur e 20. 00 ° C
•
Enter the temperature of the water to be added here. This value should
be as accurate as possible as it is used to calculate the density of the
water. Press
Di spl ay:
par amet er s
>GLP opt i ons
•
As the ‘>GLP parameters’ submenu contains no further menu items,
an automatic return is made to the submenu selection. Press
Di spl ay:
>GLP opt i ons
l i qui d: wat er
•
Press
Di spl ay:
>GLP opt i ons
wei ght f r om: bal ance
•
Use
interface and that the weight will be transmitted by the balance.
*) If the balance cannot be or is not to be connected to the RS interface then the