6.2 Error messages
711 Liquino, Instructions for use
146
Error messages relating to data transmission:
Note:
If neither a printer nor a computer is connected the report output
must be switched off.
General form of error message:
* * * RS er r or xx
Error during data transfer or printing. Check whether the connection
cable has been attached correctly and that the RS232 parameters of
the transmitting and receiving instruments are the same. See
Reception errors:
* * * RS er r or 36
Parity error
Exit:
* * * RS er r or 37
Framing error. The receiving instrument may not be ready to receive
data. Exit:
receiving instrument.
* * * RS er r or 38
Overrun error. At least 1 character could not be read.
Exit:
* * * RS er r or 39
The internal processing buffer in the Liquino is overloaded (>128
characters). Exit:
Transmission errors:
* * * RS er r or 42
CTS=OFF. Handshake not satisfied for more than 1 s.
Exit:
* * * RS er r or 43
The Liquino transmission has been interrupted by XOFF for at least 6
s. Exit:
* * * RS er r or 45
The receive buffer of the Liquino contains an incomplete character
string (L
F
is missing). The Liquino transmission is therefore blocked.
Exit: send L
F
or
Warnings
* * * Dosi ng not cont i nuous. Lower r at e * * *
You have defined a dosing rate which is too high for tandem dosing. Inter-
ruption-free dosing is not possible as a filling process (including stopcock
switching) takes longer than dosing a cylinder volume. Press
reduce the dosing rate to 85% at the most of the maximum possible dosing
rate for the dosing unit. Dosing can then be continued with
* * * Rat e t o hi gh. Adj ust par amet er ! * * *
The necessary dosing rate required to fulfill the task you have defined can-
not be reached. Press
extend the dosing period. Dosing can then be continued with
carried out at the highest possible dosing rate. However, the defined dosing
time cannot be observed in this case.