4 example of a report - cont mode – Metrohm 711 Liquino User Manual
Page 105
![background image](/manuals/587118/105/background.png)
4. Detailed description
711 Liquino, Instructions for use
99
4.6.4
Example of a report - CONT mode
Preparation of solutions and dilutions
Parameter Report (molar fraction concentration
as example.)
711 Liquino 0220 711.0011
date 1998-10-02 time 09:26
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: CONT user method: Sol-Mol2
reagent NaNO2
parameters
>CONT Parameter
type: subst.conc
target content 0.1 Mol/L
target volume 20.0 mL
factor 1.0
molar mass 69.0 g/Mol
>CONT dosing unit
dosing drive: DOS1
dos.unit: default
>CONT rates
feed rate max mL/min
fill rate max mL/min
>CONT preselections
req.ident: id1
req. weight: on
>>CONT prompts
id1 reagent
id2 Id2
id3 Id3
>CONT reports
report1: result
report2: off
report3: off
------------
← instrument number and program version
← date and time
← 3-line report header
← mode and method name
← ID1, identification with preselected display text
← dosing parameters
← dosing unit settings
← dosing and filling rate settings
← request settings
← report definitions
← identification of reproduced reports
Result Report (molar fraction concentration as
example.)
711 Liquino 0220 711.0011
date 1998-10-02 time 09:26
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: CONT user method: Sol-Mol2
reagent NaNO2
subst.conc 0.100 Mol/L
factor 1.000000
molar mass 69.0000 g
weight 0.12640 g
volume dispensed 18.319 mL
============
← see above
← targets
← sample weight
← effectively dosed-in volume
← identification for original report
Result Report (volume fraction as example)
711 Liquino 0220 711.0011
date 1998-10-02 time 09:31
*** Dev. Dept.
*** dm
*** Metrohm Ltd
mode: CONT user method: Std-ppm
standard chloride
volume fraction 25.000 ppm
target volume 100.000 ppm
volume dispensed 5.000 mL
============
← see above
← targets
← target content
← effectively dosed-in volume
← identification for original report