Metrohm 711 Liquino User Manual
Page 78

4.3 Configuration
Liquino 711, Instructions for use
72
>keyboar d opt i ons
l ock conf i gur at i on: of f
Locks
on
off
= configuration menu is not available
>keyboar d opt i ons
l ock par amet er s of f
Locks key
on
off
= Parameters cannot be altered
>keyboar d opt i ons
>>user met hods
Submenu for locking the memory functions.
Press
The settings in this submenu only affect the internal method memory. The
memory card can still be used. Memory cards can be protected against
accidental erasure by activating the write protection switch on the card
itself. Please read the information given on the sheet accompanying the
memory card.
>>user met hods
l ock met hod r ecal l : of f
Locks method loading
on
off
= no methods can be loaded from the internal
method memory
>>user met hods
l ock met hod st or e: of f
Locks method storage
on
off
= no methods can be stored in the internal meth-
od memory
>>user met hods
l ock met hod del et e: of f
Locks method delete
on
off
= no methods can be deleted in the internal
method memory
>keyboar d opt i ons
l ock di spl ay: of f
Switches off display
during operation via remote control
on
off
= the display remains locked until the Liquino is switched
off. However, it can be operated via the remote control,
see ‘Remote Control Reference'.
set up
>gr aphi cs pr i nt out
Settings for printing out curves
Navigate with
>gr aphi cs pr i nt out
r el at i ve wi dt h 0. 50
Width of the curve printout, referred to the
page format of the printer
0.3…1.00
>gr aphi cs pr i nt out
r el at i ve hei ght 0. 50
Height of the curve printout, referred to the
page formatof the printer
0.3…1.00
set up
>show/ hi de modes
Visibility of the individual modes
Navigate with