6 check display contrast, 7 check keyboard – Metrohm 711 Liquino User Manual
Page 159
![background image](/manuals/587118/159/background.png)
6. Appendix
711 Liquino, Instructions for use
153
6.5.6
Check display contrast
This diagnosis step allows the contrast setting variability function to be
checked.
•
Prepare instrument for diagnosis (see Sect. 6.5.3).
•
If necessary press key <9> several times until
di agnosi s
>di spl ay cont r ast t est
•
After pressing the
tinuously with changing contrast settings for an optical check of the
display.
>di spl ay cont r ast t est
•
The test is exited by pressing the
di agnosi s
>key t est
6.5.7
Check keyboard
This diagnosis step allows the functioning of all the keyboard keys to be
checked.
•
Prepare instrument for diagnosis (see Sect. 6.5.3).
•
If necessary press key <9> several times until
di agnosi s
>key t est
•
>key t est
•
Press all keys in sequence and observe the reaction on the display
with the help of the following keyboard matrix.
Fig. 1
Keyboard matrix
The corresponding matrix code for the pressed-down key appears in the
display. The matrix code can be seen from the keyboard matrix along-
side.
For example, when the
must appear:
>key t est
KEY: 31 CONFI G
•
The test is exited by pressing the
di agnosi s
>r emot e t est