4. Detailed description
711 Liquino, Instructions for use
115
>SMPL meas. cel l r i nsi ng
r i nsi ng vol ume 50. 0 mL
Solvent volume for rinsing the cell
0.1…250
with 'external pump off'
>SMPL meas. cel l r i nsi ng
r i ns. r at e max mL/ mi n
Dosing rate for rinsing the cell
0.001…166.66,
max
with 'external pump off'
= highest possible dosing rate (depends on type of
dosing unit), select with
par amet er s
>SMPL dosi ng uni t s
Settings for dosing units and auxiliary dosing devices
Navigate with
>SMPL dosi ng uni t s
dos. uni t 1: def aul t
Name of dosing unit used
selection list
see configuration '
>dosi ng uni t s'
>SMPL dosi ng uni t s
>>aux. sol ut . BEFORE
Submenu for definition of dosing proce-
dures before sample transfer
>>aux. sol ut . BEFORE no1
dosi ng dr i ve: of f
Dosing connection for addition of first
auxiliary solution
DOS2, DOS3, DOS4, off
>>aux. sol ut . BEFORE no1
dos. uni t : def aul t
Name of dosing unit used
selection list
see configuration '
>dosi ng uni t s'
>>aux. sol ut . BEFORE no1
vol ume 1. 000 mL
Volume of first auxiliary solution to be
added
0.0001…99999.9
>>aux. sol ut . BEFORE no1
f eed r at e max mL/ mi n
Dosing rate of auxiliary dosing device
0.001…166.66,
max
= highest possible dosing rate (depends on type of
dosing unit), select with
>>aux. sol ut . BEFORE no1
f i l l r at e max mL/ mi n
Filling rate of auxiliary dosing device
0.001…166.66,
max
= highest possible filling rate (depends on type of
dosing unit), select with
>>aux. sol ut . BEFORE no1
wai t t i me 00: 00: 00
Waiting time after addition of auxiliary
solution
HH:MM:SS
>>aux. sol ut . BEFORE no2
dosi ng dr i ve: of f
Dosing connection for addition of second
auxiliary solution
DOS2, DOS3, DOS4, off …
…up to 3 auxiliary dosing
devices