beautypg.com

Metrohm 862 Compact Titrosampler User Manual

Page 92

background image

6.7 Configuring external devices

■■■■■■■■■■■■■■■■■■■■■■

82

■■■■■■■■

862 Compact Titrosampler

Editing the COM1 settings
Under COM1 settings the interface parameters for the connected bal-
ance are set.

Baud rate

Transfer rate in characters per second.

Selection

1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200

Default value

9600

Data bits

Number of data bits.

Selection

7 | 8

Default value

8

Stop bits

Number of stop bits.

Selection

1 | 2

Default value

1

Parity

Type of parity testing.

Selection

even | none | odd

Default value

none

Handshake

Type of data transfer protocol.

Selection

hardware | software | none

Default value

hardware

Note

In case of communication problems, try the software handshake (soft-
ware
).

Editing the COM2 settings
Under COM2 settings the interface parameters for devices connected to
the RS-232/2 connector of the RS-232 USB Box are set (e.g. PC). The
parameters and input ranges are the same as for the COM1 interface.