5 Operation

862 Compact Titrosampler

■■■■■■■■

47

Sample size

Sample size. The value of the sample size can be used in calculations as
the variable C00.

Range

–999999999 ... 9999999999

Default value

1.0

Unit

Unit of sample size.

Selection

g | mg | µg | mL | µL | pieces | User-defined

Default value

g

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.

Sending the sample size from a balance

Note

In order for a balance to be able to send a sample size to the sample
table, the sample table has to be switched on in the sample series dia-
log (see "Starting a sample series", page 49).

If the sample size is sent directly from the balance, then it will always be
entered in a new line at the end of the sample table. It does not matter
which line is highlighted or whether the sample table is even opened. The
sample size is entered in line 5 in the above example.

Note

If you would like to enter the sample size in a particular line, then you
must open the corresponding editing dialog (i.e. the Sample data dia-
log is displayed).

If the editing dialog for the sample size is opened, then the sent value