10 printing a report manually – Metrohm 862 Compact Titrosampler User Manual
Page 69

■■■■■■■■■■■■■■■■■■■■■■
5 Operation
862 Compact Titrosampler
■■■■■■■■
59
The result and the sample size of each determination are shown.
On/Off
Remove the selected determination from the statistics. The line will then
be marked with an asterisk (*), the statistics will be recalculated automati-
cally. If several calculations are defined in the method, then all the results
will be removed from the statistics.
Deleting statistical data
All statistical data is deleted with the Reset function. The statistics data is
deleted automatically in following cases:
■
when all of the determinations of the determination series have been
carried out and a new determination has been started afterwards.
■
when a new method is loaded.
Adding a determination to a determination series
In case that, for example, a determination was faulty and had to be
removed from the statistics, an additional sample can be added to a deter-
mination series with the Increase function. The second number in the
Statistics line will be increased automatically by one.
5.10
Printing a report manually
Menu
▶ Print reports
To print a report manually, proceed as follows:
1 Open the main menu
■
In the main dialog, select Menu and press [OK].
2 Open the print dialog
■
Select the menu item Print reports and press [OK].
The dialog window with the available reports opens: