7.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

92

■■■■■■■■

862 Compact Titrosampler

Stop volume

The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.

Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off

Stop meas. value

The titration is stopped when the specified measured value has been
reached since the start of the titration.

Measuring mode pH:
Range

–20.000 ... 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Range

–1250.0 ... 1250.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Range

–125.0 ... 125.0 µA

Selection

off

Default value

off

Stop EP

The titration is stopped when the specified number of equivalence points
has been found.

Range

1 ... 9

Default value

9

Selection

off

Volume after EP

This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.

Range

0.01000 ... 9999.99 mL

Selection

off

Default value