Performing a sample series

Samples can be placed anywhere on the rack. They are processed in
accordance with ascending rack position.

The following is to be observed:

In addition to the sample vessels, a rinsing beaker has to be placed on
the last rack position, marked with the sign

▲. This vessel must be

either empty or filled with a rinsing solution, depending on the auto-
mation sequence, see chapter 4.4, page 29ff.

5.6.1

Starting the sample series

Starting a sample series

A suitable method must be loaded before a sample series is started (see
Chapter 5.4.3, page 43)
. The necessary parameters (see chapter 7.1,
page 85ff)
) can then be modified.

If the sample table is used and if it contains methods defined, these meth-
ods will be used. In this case, previously loading a certain method is not
necessary.

1 Define the sample series

Press the [START] key.

You can now select the quantity and the first rack position of the
samples to be processed as well as the location of the sample data.

2 Enter the number of samples

Select Number of samples and press [OK].

Enter the number of samples. Table means that all samples in the
sample table will be processed until it will be empty.

Close the entry dialog with [BACK] or Accept.

3 Enter the rack position of the first sample

Select Next sample pos. and press [OK].