7 Parameters

862 Compact Titrosampler

■■■■■■■■

133

Note

The setting of the stirring rate under Menu

Parameters Automa-

tion applies only for stirring during the immersion of the electrode in
the rinsing beaker. The stirring rate during the determination is set
under Menu

Parameters Titration parameters.

7.8

Automation: Rinsing in sample

Menu

Parameters Automation

Automation

Display of the template used for the automation sequence.

Dripping time

Waiting time after the titration head moves out of the sample beaker.

Range

0 ... 999 s

Default value

3 s

Aspiration time

Aspiration time of Pump 2, in case an 843 Pump Station is connected. It
runs before the rinsing and after the rinsing.

Range

0 ... 999 s

Default value

10 s

Rinsing time

Rinsing time of Pump 1, if an 843 Pump Station is connected. Rinsing
pump and aspiration pump run during the rinsing time.

Range

0 ... 999 s

Default value

5 s