5.11 Manual control

■■■■■■■■■■■■■■■■■■■■■■

64

■■■■■■■■

862 Compact Titrosampler

2 Select the dosing function

Select the Dosing item.

In the function bar, select DOS and press [OK].

3 Configure the dosing function

Note

The dosing and filling rate should be decreased for viscous liq-
uids.

The maximum dosing and filling rate depends on the cylinder
volume (see Chapter 10.1.1, page 139).

Enter the dosing rate.

Enter the filling rate.

4 Start dosing

Press [START].

The status changes to busy, the dosed volume is displayed. When a
cylinder volume has been added, the dosing cylinder will be refilled
automatically.

5 Fill the cylinder

Press [STOP] or [BACK].

The dosing cylinder is filled. If you start the filling with [BACK], then
the dialog will also be exited.