4 printer, 5 system initialization, Printer – Metrohm 862 Compact Titrosampler User Manual
Page 153

■■■■■■■■■■■■■■■■■■■■■■
10 Appendix
862 Compact Titrosampler
■■■■■■■■
143
Key of the 862 Compact Titro-
sampler
Mouse function
[BACK]
Right mouse button
[
⇧] [⇩] [⇦] [⇨]
Mouse movement vertical/hori-
zontal
[
⇧] [⇩]
Scroll wheel vertical
10.4.4
Printer
The range of USB printers available is extremely varied and constantly
changing. The following points must be taken into account when select-
ing a printer:
■
USB interface necessary
■
Printer language: HP-PCL, Canon BJL Commands, Epson ESC P/2 or
ESC/POS
Note
Inexpensive printers are often designed solely for use with a PC and
may not be equipped with one of the printer languages listed above.
Such models are not suitable for this reason.
10.5
System initialization
In very rare instances, it could happen that a faulty file system (e.g.
because of a program crash) will lead to an impairment of program func-
tioning. The internal file system needs to be initialized in such cases.
Caution
All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.
We recommend creating a backup copy of the system at regular inter-
vals in order to avoid data losses.
After a system initialization the program versions and language files do
not have to be reloaded. Only the selection of the dialog language may
have to be reset in the system settings.