7 Parameters
862 Compact Titrosampler
■■■■■■■■
125
Selection
on | off
Default value
off
7.3.8
Statistics
Menu
▶ Parameters ▶ Statistics
The statistics calculation of a multiple determination is switched on under
Statistics and it is defined how many determinations the series contains.
Statistics
If this funtion is switched on, then statistics calculations will be carried out
for all of the defined results.
Selection
on | off
Default value
off
Number of samples
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination must be added to the determination series,
because one determination was incorrect, for example, then this can be
accomplished in the statistical overview (see Chapter 5.9, page 58).
Range
2 ... 20
Default value
3
7.3.9
Reports
Menu
▶ Parameters ▶ Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated results, equivalence points and
endpoints, sample data, etc.
Selection
on | off
Default value
off
Curve
Selection
on | off
Default value
off