beautypg.com

3 temperature, Temperature – Metrohm 862 Compact Titrosampler User Manual

Page 158

background image

11.1 Measuring inputs

■■■■■■■■■■■■■■■■■■■■■■

148

■■■■■■■■

862 Compact Titrosampler

Measuring mode
Upol

Determination with adjustable polarization voltage.

Polarization
voltage

–1200…1200 mV (increment: 10 mV)
–1250…–1210 mV / +1210…+1250 mV: non-guaranteed values,
dependent on reference voltage +2.5 V

Measuring
range

–120…120 µA

Resolution

0.01 µA

Measuring
accuracy

11.1.3

Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.

R (25 °C) and B value can be configured for NTC sensors.

Measuring range

Pt1000

–150…250 °C

NTC

–5…250 °C
(For an NTC sensor with R (25 °C) = 30000

Ω and B (25/50) = 4100 K)

Resolution

Pt1000

0.1 °C

NTC

0.1 °C

Measuring accu-
racy

Pt1000

± 0.2 °C
(applies for measuring range –20…150 °C)

NTC

±0.6 °C
(applies for measuring range –10…40 °C)