7.3 Endpoint titrations (SET)
■■■■■■■■■■■■■■■■■■■■■■
116
■■■■■■■■
862 Compact Titrosampler
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Dosing rate
Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 139).
Range
0.01 ... 166.00 mL/min
Selection
max.
Default value
max.
Pause
Waiting time, e.g. in case of a slow response of the electrode after the
start or reaction time after the dosing of a start volume.
Range
0 ... 999999 s
Default value
0 s
Request sample ID
Selection of the sample identification that is queried in the method run.
Selection
off | ID1 | ID2 | ID1&ID2
Default value
off
Request sample size
If this parameter is switched on, then the value for the sample size will be
queried.
Selection
on | off
Default value
off
Request sample unit
If this parameter is switched on, then the unit for the sample size will be
queried.
Selection
on | off
Default value
off
Hold at request
If this parameter is switched on, then the method run will be paused dur-
ing the query. If the parameter is switched off, the titration will be started
in the background.
Selection
on | off
Default value
on