5 Operation

862 Compact Titrosampler

■■■■■■■■

61

5.11

Manual control

Menu

Manual control

The following functions are available in the manual control:

Move the sample rack (Rack position)

Move the lift (Lift position)

Dosing (Dosing)

Measure (Measure)

Stir (Stir)

The available subfunctions are listed for each function in the function bar.

5.11.1

Rotating the sample rack

If the Rack position line is selected, then the arrow keys [

] and []

can be used to select one of the following functions, which can then be
executed by pressing [OK]:

Next

The lift is moved upward and the next-higher
rack position is placed in front of the lift.

If the [OK] key remains pressed, the rack auto-
matically moves to the next position.

Previous

The lift is moved upward and the next-lower
rack position is placed in front of the lift.

If the [OK] key remains pressed, the rack auto-
matically moves to the next position.

Reset

The rack is initialized. The lift is moved upward
and the sample rack is rotated to the starting
position. At the same time, the starting position
(Next sample pos.) is reset to 1 for the start of
the next sample series.

The rack position display is always updated as soon as the rack is in the
new position.