beautypg.com

2 lift settings (lift), Lift settings (lift) – Metrohm 862 Compact Titrosampler User Manual

Page 82

background image

6.2 Lift settings (Lift)

■■■■■■■■■■■■■■■■■■■■■■

72

■■■■■■■■

862 Compact Titrosampler

Temperature sensor

The instrument supports the use of two different temperature measure-
ment techniques:

NTC (Negative Temperature Coefficient)

Pt1000 (Platinum resistance)

Select here the type that has been connected to the instrument. If an NTC
sensor is used, then it is also necessary to enter two sensor characteristics.
These characteristics are listed in the specifications of the sensor.

Selection

Pt1000 | NTC

Default value

Pt1000

R (25 °C)

This parameter is visible only when Temperature sensor = NTC.

Nominal resistance of the NTC sensor at 25 °C.

Range

1000 ... 99999 ohm

Default value

30000 ohm

B value

This parameter is visible only when Temperature sensor = NTC.

Material constant of the NTC sensor. B values of NTC sensors are fre-
quently based on different reference temperatures (usually 25 °C and
50…100 °C).

Range

1000 ... 9999 K

Default value

4100 K

6.2

Lift settings (Lift)

Menu

System Lift

Work position

The working height of the lift can be set to the desired value. This is
accomplished by means of the direct operation of the lift.