6.2 Lift settings (Lift)
■■■■■■■■■■■■■■■■■■■■■■
72
■■■■■■■■
862 Compact Titrosampler
Temperature sensor
The instrument supports the use of two different temperature measure-
ment techniques:
■
NTC (Negative Temperature Coefficient)
■
Pt1000 (Platinum resistance)
Select here the type that has been connected to the instrument. If an NTC
sensor is used, then it is also necessary to enter two sensor characteristics.
These characteristics are listed in the specifications of the sensor.
Selection
Pt1000 | NTC
Default value
Pt1000
R (25 °C)
This parameter is visible only when Temperature sensor = NTC.
Nominal resistance of the NTC sensor at 25 °C.
Range
1000 ... 99999 ohm
Default value
30000 ohm
B value
This parameter is visible only when Temperature sensor = NTC.
Material constant of the NTC sensor. B values of NTC sensors are fre-
quently based on different reference temperatures (usually 25 °C and
50…100 °C).
Range
1000 ... 9999 K
Default value
4100 K
6.2
Lift settings (Lift)
Menu
▶ System ▶ Lift
Work position
The working height of the lift can be set to the desired value. This is
accomplished by means of the direct operation of the lift.