7.5 Automation: Dipping in special

■■■■■■■■■■■■■■■■■■■■■■

130

■■■■■■■■

862 Compact Titrosampler

Parameters

All of the parameters of the current method are printed out with the para-
meter report.

Selection

on | off

Default value

off

PC/LIMS

The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 80)
.

The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt
.

Selection

on | off

Default value

off

7.5

Automation: Dipping in special

Menu

Parameters Automation

Automation

Display of the template used for the automation sequence.

Dripping time

Waiting time after the titration head moves out of the sample beaker and
out of the rinsing beaker.

Range

0 ... 999 s

Default value

3 s

Rinsing time

Waiting time during which the electrode remains immersed in the rinsing
beaker.

Range

0 ... 999 s

Default value

5 s

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 U/min. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 140. The optimum stirring
rate can be tested in the manual control.