7.5 Automation: Dipping in special
■■■■■■■■■■■■■■■■■■■■■■
130
■■■■■■■■
862 Compact Titrosampler
Parameters
All of the parameters of the current method are printed out with the para-
meter report.
Selection
on | off
Default value
off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 80).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection
on | off
Default value
off
7.5
Automation: Dipping in special
Menu
▶ Parameters ▶ Automation
Automation
Display of the template used for the automation sequence.
Dripping time
Waiting time after the titration head moves out of the sample beaker and
out of the rinsing beaker.
Range
0 ... 999 s
Default value
3 s
Rinsing time
Waiting time during which the electrode remains immersed in the rinsing
beaker.
Range
0 ... 999 s
Default value
5 s
Stirring rate
Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 U/min. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 140. The optimum stirring
rate can be tested in the manual control.