7 Parameters

862 Compact Titrosampler

■■■■■■■■

129

Number of buffers

Number of buffers that are used for calibration. If calibration is accom-
plished with more than two buffers, then they can be used repeatedly
give them more statistical weight. The first two buffers must however
always be different from one another.

Selection

1 | 2 | 3 | 4 | 5

Default value

2

Buffer 1 pH

This parameter is visible only when Buffer type = Special.

Range

–20.000 ... 20.000

Default value

7.000

Buffer 2 pH

This parameter is visible only when Buffer type = Special.

Range

–20.000 ... 20.000

Default value

4.000

Selection

off

Buffer 3 pH

This parameter is visible only when Buffer type = Special.

Range

–20.000 ... 20.000

Selection

off

Default value

off

Buffer 4 pH

See Buffer 3 pH.

Buffer 5 pH

See Buffer 3 pH.

7.4.3

Reports

Menu

Parameters Reports

The reports that will be printed out automatically in connection with a cal-
ibration are defined under Reports.

Results

The result report contains the specifications for the calibration (slope,
pH(0), etc.).

Selection

on | off

Default value

off