beautypg.com

3 managing sensors, 1 general, Managing sensors – Metrohm 862 Compact Titrosampler User Manual

Page 83: General

background image

■■■■■■■■■■■■■■■■■■■■■■

6 System settings

862 Compact Titrosampler

■■■■■■■■

73

Three functions can be selected from the function bar with [

] and []

and then run by pressing [OK]:

Work pos. moves the lift to the current working height.

Up moves the lift 6 mm upward.

Down moves the lift 6 mm downward.

When this dialog page is exited, the respective current lift position will be
applied as Work pos..

Range

0 ... 132 mm (Increment: 6)

Default value

60 mm

Initial lift pos.

After the 862 Compact Titrosampler has been switched on, the lift moves
all the way to the top to the home position for the initialization of the
drive. It can then be moved back down to the working height if desired.

Selection

shift pos. | work pos.

Default value

shift pos.

shift pos.
Resting position (0 mm) all the way up
work pos.
The set working height

6.3

Managing sensors

6.3.1

General

Menu

System Sensors

Three standard sensors are defined in the sensor list: pH electrode,
Metal electrode and Temperature sensor. These sensors cannot be
deleted or renamed. The sensor list can contain a maximum of 10 sensors.

Every sensor is identified with an unambiguous name. This means that it is
not possible to use the same name twice, e.g. for a pH electrode and for a
metal electrode.

Edit

Edit the data of the selected sensor, see following chapter.