5 Operation
862 Compact Titrosampler
■■■■■■■■
53
Note
Enter the rack position of the first buffer solution as the Next sample
pos. at the start of the sample series.
When specifying the Number of samples, enter (for the calibration)
one sample more than is to be titrated. The calibration is considered to
be one single sample processing, even if several buffer solutions are
being processed. The number of lines in the sample table is the deter-
mining factor. You can also select Table as the Number of samples.
5.7
Live modifications
5.7.1
Editing the sample data of the running determination
The sample data can be entered or modified in the main dialog while a
determination is running. In calculations always the sample data entered
at the end of the titration in the main dialog is used.
Proceed as follows to edit the sample data:
1 Display the main dialog
■
Press [BACK].
The main dialog is displayed. The determination continues to run in
the background.
2 Edit the sample data
■
Edit the sample data and accept with Accept or [BACK].
3 Display the live dialog
■
Press [BACK].
or
■
Select Menu and press [OK].
■
Select the menu item Live dialog and press [OK].
The live dialog is displayed once again.