7 configuring external devices, Configuring external devices, Figure 29 – Metrohm 862 Compact Titrosampler User Manual
Page 90: Directory structure on the usb stick

6.7 Configuring external devices
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
862 Compact Titrosampler
Figure 29
Directory structure on the USB stick
Backup
All of the files of the backup copy are stored in
this directory. The directory will be created the
first time a backup copy is saved.
Files
Exported methods will be stored in this directory.
The directory will be created the first time a
method is exported.
Only methods being located in this directory can
be imported.
pc_lims_report
PC/LIMS reports are stored in this directory as a
TXT file. The directory will be created the first
time a PC/LIMS report is printed out.
6.7
Configuring external devices
Menu
▶ System ▶ External devices
PC/LIMS report
Specification of the memory location for the PC/LIMS report. The PC/LIMS
report is a machine-readable report with all of the data important for a
determination. It can be saved as follows:
■
as a TXT file on a USB stick.
■
to a LIMS via an RS-232 interface. The 6.2148.030 RS-232/USB Box is
required for this purpose.
Selection
COM2 | USB Stick
Default value
USB Stick
COM2
The report is sent via the serial COM2 interface. The interface parame-
ters set in the dialog COM2 settings are used (see "Editing the COM2
settings", page 82).
USB Stick
The report will be saved as a TXT file on the USB stick in the folder
pc_lims_report.
Printer
If a printer is connected, then the printer type needs to be defined here in
order for the reports to be printed out correctly.
The printers that have the designation ESC-POS are so-called POS printers
(point-of-sale printers), i.e. they print on continuous paper.