9.2 Miscellaneous
■■■■■■■■■■■■■■■■■■■■■■
138
■■■■■■■■
862 Compact Titrosampler
Problem
Cause
Remedy
The electrode responds too
slowly.
Replace the electrode.
9.2
Miscellaneous
Problem
Cause
Remedy
No report is printed.
The printer is not recog-
nized by the instrument.
■
Switch the 862 Compact Titrosampler off
and on again.
■
Use a hub on the USB connector and con-
nect the printer to the hub.
The printer model is not
compatible.
Use a printer which fulfills the required specifi-
cations (see Chapter 10.4.4, page 143).
The USB keyboard
or the PC mouse
does not function.
The keyboard or mouse is
not recognized by the
instrument.
■
Switch the 862 Compact Titrosampler off
and on again.
■
Use a hub on the USB connector and con-
nect the keyboard or mouse to the hub.
The keyboard or the mouse
is not compatible.
Use a model which fulfills the required specifi-
cations (see Chapter 10.4, page 141).
The display is no
longer readable.
The contrast is set incor-
rectly.
Adjust the contrast correctly (see Chapter 6.1,
page 69).
Mettler XP balances
send "R" or "O ----"
as ID1.
The automatic calibration
of the balance is switched
on.
Switch off the automatic calibration.
Message 020-511
"Action not possi-
ble" is displayed.
The USB stick is no longer
connected.
1. Connect the USB stick.
2. Switch the instrument off and on again.
The USB stick is full.
■
Use a different USB stick.
■
Delete files with the aid of a PC.