7 Parameters
862 Compact Titrosampler
■■■■■■■■
115
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Cond. stop volume
Maximum permissible volume that can be dosed during conditioning.
Conditioning is stopped when the specified volume is dosed. If condition-
ing is continued by pressing [START] once again, then the titrant volume
that has already been dosed will not be taken into account; i.e. the dosing
starts again at zero. The stop volume should be adjusted to the size of the
titration cell in order to prevent any overflow.
Range
0.00000 ... 9999.99 mL
Default value
20.0000 mL
Selection
off
Cond. stop time
Maximum permissible time over which conditioning may take place. Con-
ditioning is stopped when the specified time has elapsed.
Range
0 ... 999999 s
Selection
off
Default value
off
7.3.2
Start conditions
Menu
▶ Parameters ▶ Start conditions
The parameters that are carried out before the start of titration are
defined under Start conditions.
Activation pulse
Output of an activation pulse on a remote line. This activation pulse starts
a connected Dosimat.
Selection
on | off
Default value
off
Start delay time
Waiting time after the start of the determination, before titration takes
place. An auxiliary solution can, e.g. be added with a Dosimat during this
time (parameterization at the Dosimat). The Activation pulse parameter
must however be switched on for this purpose.
Range
0 ... 999999 s
Default value
0 s
Start volume
Volume that is dosed prior to the start of the titration.