7.1 Dynamic equivalence point titrations (DET)

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

862 Compact Titrosampler

Upper limit

This parameter is visible only when Window = on.

Measured value for the upper limit.

Measuring mode pH:
Range

–20.000 ... 20.000

Default value

20.000

Measuring mode U, Ipol:
Range

–1250.0 ... 1250.0 mV

Default value

1250.0 mV

Measuring mode Upol:
Range

–125.00 ... 125.00 µA

Default value

125.00 µA

EP criterion

The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.

Range

0 ... 200

Default value

5

EP recognition

This parameter allows you to filter out only the equivalence points that are
being sought.

for Window = off
Selection

all | greatest | last | off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
off
No evaluation takes place.

for Window = on
Selection

first | greatest | last

Default value

first