5 managing common variables, 1 general, Managing common variables – Metrohm 862 Compact Titrosampler User Manual
Page 87: General
■■■■■■■■■■■■■■■■■■■■■■
6 System settings
862 Compact Titrosampler
■■■■■■■■
77
Range
–999999999 ... 9999999999
Default value
1.000
Titer unit
Unit of the titer.
Selection
µmol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
µg/L | ppm | % | mEq/L | empty | User-defined
Default value
empty
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Date titer det.
Date of the last titer determination.
Monitoring
Switching the titer monitoring on and off.
Selection
on | off
Default value
off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Range
1 ... 999 d
Default value
999 d
6.5
Managing common variables
6.5.1
General
Menu
▶ System ▶ Common variables
The instrument offers the possibility of saving five method-independent
variables, so-called common variables. These variables remain saved in
the instrument and can be used in future calculations. Common variables
are useful, e.g. for the following applications:
■
Determination of a blank value which will be taken into account during
the content determination of the sample.
■
Determination of the content of a standard solution, which will be
taken into account during the content determination of the sample.