4 control parameters ep1, Control parameters ep1 – Metrohm 862 Compact Titrosampler User Manual
Page 129
■■■■■■■■■■■■■■■■■■■■■■
7 Parameters
862 Compact Titrosampler
■■■■■■■■
119
+
Positive measured value alteration, i.e. in the direction of a higher pH
value, greater voltage or greater current.
–
Negative measured value alteration, i.e. in the direction of a lower pH
value, lesser voltage or lesser current.
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.
Extraction time
Minimum duration of the titration. Titration will not be stopped during the
extraction time, even if the endpoint has already been reached. Titration is
however stopped if a stop condition is fulfilled during this time (see Chap-
ter 7.3.6, page 122). The entry of an extraction time may be advisable,
e.g. for the titration of sparingly soluble samples.
Range
0 ... 999999 s
Default value
0 s
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Range
0.1 ... 999999.0 s
Default value
2.0 s
7.3.4
Control parameters EP1
Menu
▶ Parameters ▶ Control parameters EP1
The control parameters for the first endpoint are defined under Control
parameters EP1.
Endpoint 1 at
Measured value for the first endpoint.
Measuring mode pH:
Range
–20.000 ... 20.000
Selection
off
Default value
off
Measuring mode U and Ipol:
Range
–1250.0 ... 1250.0 mV
Selection
off
Default value
off