6 System settings
862 Compact Titrosampler
■■■■■■■■
79
The assignment of the result to a common variable occurs automatically in
accordance with the following scheme:
■
Result R1
⇨ Common variable CV01
■
Result R2
⇨ Common variable CV02
■
etc.
6.6
File management
Menu
▶ System ▶ File management
Note
This menu item is visible only when a USB stick has been connected as
an external storage medium.
Methods can be imported and deleted from a USB stick in this dialog.
Only methods located in the Files directory are displayed in the list (see
"Directory structure on the USB stick", page 79).
A backup copy can be made of the system (all data and settings). Similarly,
an existing backup copy can be stored.
Import
Import the selected method.
Delete
Delete the selected method.
Backup
Create a backup copy of all data and settings on the USB stick.
Note
Only one backup copy can be created on the same USB stick.
If a backup copy has already been stored on the stick, then it will be
overwritten when this function is carried out again.
Restore
Load the backup copy from a connected USB stick.
Directory structure on the USB stick
A directory with the instrument number is generated on the USB stick. The
structure within the directory appears as follows: