7.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
106
■■■■■■■■
862 Compact Titrosampler
7.2.3
Stop conditions
Menu
▶ Parameters ▶ Stop conditions
The conditions for stopping a titration are defined under Stop condi-
tions.
Stop volume
The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off
Stop meas. value
The titration is stopped when the specified measured value has been
reached since the start of the titration.
Measuring mode pH:
Range
–20.000 ... 20.000
Selection
off
Default value
off
Measuring mode U, Ipol:
Range
–1250.0 ... 1250.0 mV
Selection
off
Default value
off
Measuring mode Upol:
Range
–125.0 ... 125.0 µA
Selection
off
Default value
off
Stop EP
The titration is stopped when the specified number of equivalence points
has been found.
Range
1 ... 9
Default value
9
Selection
off
Volume after EP
This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.