7.3 Endpoint titrations (SET)

■■■■■■■■■■■■■■■■■■■■■■

124

■■■■■■■■

862 Compact Titrosampler

7.3.7.2

Editing a calculation

Result name

The result name is the text which will be shown in the result view and in
the report.

Input

12 characters

Default value

empty

R1=…R5=

Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 5.3, page 39).

Input

44 characters

Default value

empty

Decimal places

Number of decimal places used to display the result.

Range

0 ... 5

Default value

2

Result unit

The result unit is displayed and saved together with the result.

Selection

% | mol/L | mmol/L | g/L | mg/L | mg/mL | ppm |
g | mg | mL | mg/piece | °C | µL | mL/min | User-
defined

Default value

%

User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.

Save as titer

The result can be saved as titer for this selected solution. If statistics has
been switched on, then the current mean value of the determination ser-
ies will be saved.

Selection

on | off

Default value

off

Save as CV

The calculated result can be saved as a method-independent variable, a
so-called common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.