7.7 Automation: Double dipping

■■■■■■■■■■■■■■■■■■■■■■

132

■■■■■■■■

862 Compact Titrosampler

"–": clockwise rotation

Range

–15 ... 15

Default value

8

Note

The setting of the stirring rate under Menu

Parameters Automa-

tion applies only for stirring during the immersion of the electrode in
the rinsing beaker. The stirring rate during the determination is set
under Menu

Parameters Titration parameters.

7.7

Automation: Double dipping

Menu

Parameters Automation

Automation

Display of the template used for the automation sequence.

Dripping time

Waiting time after the titration head moves out of the sample beaker and
out of the rinsing beaker.

Range

0 ... 999 s

Default value

3 s

Rinsing time

Waiting time during which the electrode remains immersed in the rinsing
beaker.

Range

0 ... 999 s

Default value

5 s

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 U/min. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 140. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Range

–15 ... 15

Default value

8