5 Operation

862 Compact Titrosampler

■■■■■■■■

57

The calculated result of the last determination and details concerning the
endpoint are shown in the overview.

Curve

Display the curve of the current determination.

Recalc

Recalculate the current determination. The procedure will be executed
immediately.

Statistics

Display the statistical overview of a determination series (see Chapter 5.9,
page 58)
.

Displaying the curve
The curve of the current determination can be displayed with the Curve
function.

The arrow keys [

] and [] can be used to move to the individual measur-

ing points. A cross hair is used to show the current position on the curve.
The data (volume, measured value, time, etc.) for the respective measur-
ing point is indicated on the right-hand side.

Recalculating

Note

Recalculation cannot be undone.

All of the results are recalculated with the Recalc function. This is neces-
sary if, for example, the calculation, the titer or the sample size has been
modified.