4 Titration and automation sequences

862 Compact Titrosampler

■■■■■■■■

31

Wait for Dripping time to be completed

Move to rack position 11

Lower the lift to work position and switch on the stirrer

Wait for Rinsing time to be completed

Switch off the stirrer and move the lift upward

Wait for Dripping time to be completed

After the last sample:

Move to special beaker position

Lower the lift in the rinsing beaker to work position

4.4.3

Double dipping

This automation sequence is suitable for simple determinations.

No pump is required for rinsing or for aspiration of the sample vessels. The
titration head with electrode and burette tips is immersed in the filled rins-
ing beaker on rack position 11 and on the special beaker position after
each determination. At the same time, the rinsing solution is stirred.

If required, auxiliary solution can be added prior to the beginning of a
determination with a Dosimat/Dosimat plus.

A filled rinsing beaker should be placed on rack position 11 and on the
special beaker position.

The individual steps:

Move to the sample

Lower the lift to work position

Start the determination

If necessary, initiate dosing (Activation pulse) and switch on the stir-
rer

If necessary, wait for Start delay time

Wait for end of determination

Switch off the stirrer and move the lift upward

Wait for Dripping time to be completed

Move to rack position 11

Lower the lift to work position and switch on the stirrer

Wait for Rinsing time to be completed

Switch off the stirrer and move the lift upward

Wait for Dripping time to be completed

Moving to special beaker position

Lower the lift to work position and switch on the stirrer

Wait for Rinsing time to be completed

Switch off the stirrer and move the lift upward

Wait for Dripping time to be completed

After the last sample:

Lower the lift in the rinsing beaker to work position