7.1 Dynamic equivalence point titrations (DET)
■■■■■■■■■■■■■■■■■■■■■■
90
■■■■■■■■
862 Compact Titrosampler
Titration rate
slow
optimal
fast
Signal drift
– pH, U and Ipol
– Upol
20.0 mV/min
20.0 µA/min
50.0 mV/min
50.0 µA/min
80.0 mV/min
80.0 µA/min
Min. waiting time
0 s
0 s
0 s
Max. waiting time
38 s
26 s
21 s
Temperature
Manually entered titration temperature. If a temperature sensor is connec-
ted then the temperature will be measured continuously. This value is
used for temperature correction in pH measurements.
Range
–20.0 ... 150.0 °C
Default value
25.0 °C
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System
▶ Sensors.
Selection
Selection of configured sensors
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. This ensures that the correct data (titer, concentra-
tion, etc.) is always used for the calculation. Solutions are defined under
System
▶ Solutions.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct solution has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. The
validity of the titer is checked for the selected solution at the start of the
determination.
Selection
Selection of configured solution | not defined
Default value
not defined
not defined
No check takes place.
I(pol)
The polarization current is the current that is applied at a polarizable elec-
trode during the voltametric measurement. This parameter is available
only with I(pol) determinations.