6.3 Managing sensors

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

862 Compact Titrosampler

New

Add a new sensor to the list, see following chapter.

The following sensor types can be selected:

pH electrode

Metal electrode

Temperature sensor

Other sensor, e.g. Spectrosense

Delete

Delete the selected sensor from the list.

6.3.2

Editing the sensor data

Name

The designation of the sensor is used for unambiguous identification.

Input

max. 24 characters

Default value

empty

Type

The sensor type is displayed.

Slope

This parameter only appears for pH electrodes.

Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0 % is used as the slope.

Range

–999.9 ... 999.9 %

Default value

100.0 %

pH(0)

This parameter only appears for pH electrodes.

pH value of the pH electrode at 0 mV. Apart from the slope, pH (0) is the
second characteristic of the calibration curve.

Range

–20.000 ... 20.000

Default value

7.000

Calibration temp.

This parameter only appears for pH electrodes.

Temperature at which the last calibration was carried out.

Range

–20.0 ... 150.0 °C

Default value

25.0 °C