2 editing the sensor data, Editing the sensor data – Metrohm 862 Compact Titrosampler User Manual
Page 84
![background image](/manuals/586771/84/background.png)
6.3 Managing sensors
■■■■■■■■■■■■■■■■■■■■■■
74
■■■■■■■■
862 Compact Titrosampler
New
Add a new sensor to the list, see following chapter.
The following sensor types can be selected:
■
pH electrode
■
Metal electrode
■
Temperature sensor
■
Other sensor, e.g. Spectrosense
Delete
Delete the selected sensor from the list.
6.3.2
Editing the sensor data
Name
The designation of the sensor is used for unambiguous identification.
Input
max. 24 characters
Default value
empty
Type
The sensor type is displayed.
Slope
This parameter only appears for pH electrodes.
Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0 % is used as the slope.
Range
–999.9 ... 999.9 %
Default value
100.0 %
pH(0)
This parameter only appears for pH electrodes.
pH value of the pH electrode at 0 mV. Apart from the slope, pH (0) is the
second characteristic of the calibration curve.
Range
–20.000 ... 20.000
Default value
7.000
Calibration temp.
This parameter only appears for pH electrodes.
Temperature at which the last calibration was carried out.
Range
–20.0 ... 150.0 °C
Default value
25.0 °C