Operation – Leica Biosystems ASP6025 User Manual
Page 99

97
Leica ASP6025
Reagent warning messages
If the RMS is active, warning messages are output if reagents are used for too
long.
The following indicates that the warning threshold values have been exceeded:
• Exclamation mark on the station system or RTU-bottle in the SMART screen
(
Fig. 13
5).
The RMS system – adjusting warning threshold values
(continued)
• When a warning threshold value is exceeded, a warning message is also
displayed at the end of a program (after the cleaning) (
Fig. 13
7).
From there, you can press YES to go directly to the REAGENT STATUS window.
Fig. 135
Fig. 136
Fig. 137
Fig. 138
• A warning message when starting a protocol (
Fig. 13
6).
The operator is notified that the reagent has run out and can replace this
reagent before the program starts.
However, it is also possible to start the program despite this.
• The reagent status highlights the exceeded values in red in the table (
Fig. 13
8).
Displayed if:
• The number of processed cassettes is exceeded.
• The number of specified cycles is exceeded.
• The days until changing the reagent are exceeded.